STRUCTURAL AND MECHANICAL ASSESSMENT OF DENTAL IMPLANTS BASED ON TiO2 and ZrO2
Yulia Vadimovna Maskurova1, Vladislav Anatolyevich Kokoev2, Kurban Aigumovich Gusengadzhiev3, Alvina Albertovna Megrikyan4, Kamilla Medzhidovna Alieva3, Nikolay Georgievich Serov4
1Department of Therapeutic, Surgical and Pediatric Dentistry, Dental Faculty of North Ossetian State University named after K. L. Khetagurov, Vladikavkaz, Republic of North Ossetia-Alania, Russia.
2Department of Dentistry No.2, Dental Faculty of North Ossetian State Medical Academy, Vladikavkaz, Republic of North Ossetia-Alania, Russia.
3Department of Therapy, Dental Faculty of Northwestern State Medical University named after I. I. Mechnikov, St. Petersburg, Russia.
4Department of Therapy, Dental Faculty of Medical Academy named after S.I. Georgievsky, Crimean Federal University named after V.I. Vernadsky, Simferopol, Crimea, Russia.
ABSTRACT
At the stages of choosing a design and during the installation of a dental implant, its optimal parameters are taken into account. The determination of these parameters depends on the volume of bone tissue of the lower jaw and its mechanical properties. When choosing a design, it is necessary to assess the level of violations developing in the process of functioning. This article discusses two types of systems TiO2 implants and structures made of ZrO2. The possibility of using implants based on ZrO2 is evaluated by analyzing the stress-strain state of both the implant itself and the implant–bone zone, where large stress differences occur. This property makes it possible to ensure better compression performance of the implant during biting. In this task, the implant mainly works on bending, and the less its bending is, the less impact it will have on the bone tissue in the cervical region. It can be concluded that an implant made of ZrO2 causes less stress than an implant made of zirconium alloy under lateral loading during chewing.
Key words: Dental implant, Titanium, Zirconium, Oxides, Mechanical properties.
Introduction
When even one tooth is lost, functional stresses are redistributed in the maxillary system – teeth are displaced to the area of the formed defect, and the periodontal of the remaining teeth is overloaded [1]. The expediency of replacing defects in the dentition is beyond doubt. In recent years, the interest of specialists in the use of the dental implantation method in a wide dental practice has been increasing [2].
Dental implants have proven to be extremely reliable artificial intraosseous supports of dentures. Their demand in modern dentistry is because only dental implants create the possibility of replacing extended defects of dentition with fixed prostheses [3, 4]. The history of successful clinical use of implants is half a century and is largely explained by the biocompatibility, technological processing, and strength of the main structural material of implants [5]. At the same time, one of the main aspects at the stages of the patient's examination is to determine the conditions for the installation of an intraosseous implant, in particular, the availability of sufficient bone tissue and the choice of the optimal implant design [6-8].
Titanium implants are most widely represented on the Russian market, but recently zirconium dioxide structures have been used [9]. To justify the use of new implant designs, it is necessary to make sure that they are resistant to loads formed during the chewing function [10]. Both TiO2 and ZrO2 have high strength, hydrophobicity, biocompatibility, and low toxicity [11-13]. The chemical composition of the alloy determines its mechanical properties. Zirconium alloys belong to the group of alloys with solid-solution hardening and differ from intermetallic alloys, i.e. those prone to magnetization, including titanium, by high fatigue endurance characteristics that do not depend much on the structure of the metal [14]. The low structural sensitivity of zirconium makes it possible to expand the range of technological effects. It should be noted that zirconium materials tend to self-heal surface defects and are highly resistant to cracking [15]. Zirconium alloys are favorably distinguished from titanium alloys by the absence of hydrogen absorption and a tendency to hydrogen embrittlement at temperatures above 50-70 oC in the air during processing [16]. The combination of high corrosion resistance in organic compounds, high processability, crack resistance, and fatigue endurance make zirconium alloys very promising for use in the production of endoprostheses [17].
The purpose of the work was to evaluate the ability of ZrO2 implants to withstand the loads formed during operation and compare them with similar stresses that occur when using a TiO2 implant.
Materials and Methods
In this paper, we consider the installation of an implant instead of a lost incisor. The use of TiO2 and ZrO2 implants is not only because they are durable, but also bioinert materials. The installation of implants in the incisor area is because the implant must withstand the loads that fall on it during chewing, as well as being aesthetic since the front teeth are exposed when smiling [18]. At the moment, dentists prefer to install implants made of ZrO2, which has a white color in contrast to gray metals.
In this work, a model of the implant and the surrounding bone tissue is constructed for a three-dimensional situation. Consider the case when there is no central incisor on the lower jaw. Figure 1a shows a model of an implant with a smooth surface installed in the bone tissue of the lower jaw, in which the cortical (dense) and trabecular (spongy) layers are distinguished. The area of bone tissue and the implant is modeled as an area with different mechanical properties. The height of the lower jaw consists of two values: the average height of the lower jaw body (the distance from the lower edge of the jaw to the tops of the roots) is 18 mm and the height of the alveolar process is 11.5 mm [19]. The average length of the incisor is 21 mm (19-23 mm) [20], although in other sources it varies from 18.5 to 26.6 mm, while the crown height is 7.3–12.6 mm, the root height is 9.4–18.1 mm [21].
Table 1 shows the mechanical properties of TiO2 and ZrO2, as well as cortical and trabecular bone tissue. A full description of the properties and composition of an implant based on ZrO2 is presented in the international standard ISO 13356, which regulates the limit values of stresses that an implant made of this material must withstand.
Figure 1b shows the boundary conditions:
To apply the load on the implant, we will place a crown in the form of a prism on its extra-costal part, the dimensions of which correspond to the dimensions of the incisor crown. The medial-distal size of the crown between the contact points range from 4.6 to 8.2 mm, and the neck – from 3.0 to 4.9 mm. The size of the crown in the vestibular-lingual direction in the equator area is from 5.2 to 7.4 mm, and in the neck area – from 4.3 to 6.8 mm.
|
a) |
|
b) |
Figure 1. Implant diagram: a) a model of an implant with a smooth surface with a crown installed on it and the surrounding bone tissue of the lower jaw, b) boundary conditions |
Table 1. Mechanical characteristics of TiO2, ZrO2, ceramics, cortical and trabecular bone tissue
Material |
Modulus of elasticity (E), MPa |
Poisson's Ratio (ν) |
TiO2 |
110·103 |
0,3 |
ZrO2 |
210·103 |
0,3 |
Ceramic |
170·103 |
0,3 |
Cortical bone tissue |
13,7·103 |
0,3 |
Trabecular bone tissue |
6,89·103 |
0,3 |
Results and Discussion
The problem of the stress-strain state of the implant in the bone tissue was solved within the framework of the theory of elasticity in the Ansys software package [24]. In the course of solving the problem, we will compare the stresses obtained with the strength limits of TiO2, ZrO2, and bone tissue, presented in Table 2. The strength limit of the trabecular bone tissue is assumed to be 10 times less than the corresponding strength limit of the cortical bone tissue.
Figure 2a shows the stress distributions along the X-axis and the Mises stress for implants made of ZrO2. For comparison, we calculate the same problem for the case when the implant is made of TiO2 (Figure 2b). Analyzing the stress values obtained (Figure 2), we conclude that the stresses do not exceed the strength limits for both the materials from which the implants are made and for bone tissue (Table 2).
It should be mentioned that in the process of chewing, the implant is subjected not to a single load, but to cyclic loading since chewing food is because the lower jaw of a person performs vertical, horizontal, and transversal movements.
Table 2. Strength limits of TiO2, ZrO2, ceramics, cortical and trabecular bone tissue
Material |
Tensile strength, MPa |
TiO2 |
200–400 |
ZrO2 |
1400 |
Ceramic |
250 |
Cortical bone tissue |
130 |
Trabecular bone tissue |
13 |
|
a) |
|
b) |
|
c) |
|
d) |
Figure 2. Mechanical properties of implants based on TiO2 and ZrO2: stress distribution along the X-axis for an implant based on ZrO2 (a) and TiO2 (b) and the intensity of stress by Mises for an implant based on ZrO2 (c) and TiO2 (d) |
The international standard ISO 13356 recommends the use of implants with fatigue stress limits exceeding 300 MPa. The fatigue stress limit for brittle materials is equal to:
τ-1≈0.8 σ-1 |
(1) |
Where σ-1 – tensile strength.
Thus, in this work, ZrO2 with a tensile strength of 1400 MPa was considered, therefore, the fatigue stress limit for this material is 1120 MPa. Comparing the obtained stresses with this limit, we obtain that ZrO2 fully meets the strength requirements of the international standard for implants [25-27].
Conclusion
This article presents a comparative analysis of two types of dental implants made of titanium alloy and zirconium dioxide. The results of the study allow us to conclude that it is advisable to use implants made of zirconium dioxide in clinical practice. Compared with the implant, made of TiO2 it showed the best results: stress in bone tissue in the area of bone – implant below (σi=178 MPa , 13% σ-1
) than for the case of the implant from ZrO2 (σi=219 MPa
, 73% σ-1
). ZrO2 implant is more similar in mechanical properties to bone tissue, but the relative elongation of metal alloys is an order of magnitude greater than for bone tissue. This property makes it possible to ensure better compression performance of the implant during biting. In this task, the implant mainly works on bending, and the less its bending is, the less impact it will have on the bone tissue in the cervical region [28, 29]. It can be concluded that an implant made of ZrO2 causes less stress than an implant made of zirconium alloy under lateral loading during chewing.
Acknowledgments: None
Conflict of interest: None
Financial support: None
Ethics statement: None
1. Galabueva AI, Biragova AK, Kotsoyeva GA, Borukayeva ZK, Yesiev RK, Dzgoeva ZG. Optimization of modern methods of treating chronic generalized periodontitis of mild severity. Pharmacophore. 2020;11(1):47-51.
2. Remizova AA, Sakaeva ZU, Dzgoeva ZG, Rayushkin II, Tingaeva YI, Povetkin SN, et al. The role of oral hygiene in the effectiveness of prosthetics on dental implants. Ann Dent Spec. 2021;9(1):39-46. doi:10.51847/HuTuWdD0mB
3. Kotsoeva GA, Esiev RK, Toboev GV, Zakaeva RS, Kulova AA, Tsokova LV. Phytoadaptogenic cocktail use “Biorithm-E” in the complex treatment of odontogenic inflammatory diseases of the maxillofacial region. Ann Dent Spec. 2021;9(2):52-7.
4. Povetkin SN, Khasarov KA, Mazaev DS, Alishlalov SA, Kozaeva FV, Khanbetirova ZZ, et al. To the Question about the Development of Composition and Technology of Soft Medicinal Forms Ointments for the Treatment of Periodontal Disease. Ann Med Health Sci Res. 2021;11:1599-601.
5. Elani HW, Starr JR, Da Silva JD, Gallucci GO. Trends in Dental Implant Use in the U.S., 1999-2016, and Projections to 2026. J Dent Res. 2018;97(13):1424-30. doi:10.1177/0022034518792567
6. Maisigov JB, Kuznetsova GV, Magomedov AM, Adzhigova FZ, Magomedova AS, Burdukova SA, et al. Anthropometric Analysis of Digital Models of the Dentition Using 3D Technologies in Orthodontics. J Pharm Res Int. 2021;33(39B):101-5. doi:10.9734/jpri/2021/v33i40A32225
7. Yusupova MI, Mantikova KA, Kodzokova MA, Mishvelov AE, Paschenko AI, Ashurova ZAK, et al. Study of the possibilities of using augmented reality in dentistry. Ann Dent Spec. 2021;9(2):17-21. doi:10.51847/BG1ZazqXRc
8. Hight GY, Mishvelov AE, Nuzhnaya CV. New image modeling features for planning surgical interventions. Res J Pharm Biol Chem Sci. 2019;10(1):140-3.
9. Nicholson J. Titanium Alloys for Dental Implants: A Review. Prosthesis. 2020;2(2):100-16. doi:10.3390/prosthesis2020011
10. Sadyrin E, Swain M, Mitrin B, Rzhepakovsky I, Nikolaev A, Irkha V, et al. Characterization of Enamel and Dentine about a White Spot Lesion: Mechanical Properties, Mineral Density, Microstructure, and Molecular Composition. Nanomaterials (Basel). 2020;10(9):1889. doi:10.3390/nano10091889
11. Lyashenko EN, Uzbekova LD, Polovinkina VV, Dorofeeva AK, Ibragimov S-US-u, Tatamov AA, et al. Study of the Embryonic Toxicity of TiO2 and ZrO2 Nanoparticles. Micromachines. 2023;14(2):363. doi:10.3390/mi14020363
12. Yan D, Xu G, Yao Z, Liu H, Chen Y. Synergistic Effect of Sic Particles and Whiskers on the Microstructures and Mechanical Properties of Ti(C, N)-Based Cermets. Materials. 2022;15(6):2080. doi:10.3390/ma15062080
13. Blinov AV, Nagdalian AA, Arefeva LP, Varavka VN, Kudryakov OV, Gvozdenko AA, et al. Nanoscale Composite Protective Preparation for Cars Paint and Varnish Coatings. Coatings. 2022;12(9):1267. doi:10.3390/coatings12091267
14. George EP, Raabe D, Ritchie RO. High-entropy alloys. Nat Rev Mater. 2019;4(8):515-34. doi:10.1038/s41578-019-0121-4
15. Ewais OH, Al Abbassy F, Ghoneim MM, Aboushelib MN. Novel zirconia surface treatments for enhanced osseointegration: laboratory characterization. Int J Dent. 2014;2014:203940. doi:10.1155/2014/203940
16. Pozio A, Aurora A, Della Seta L, Santucci A, Tosti S. Titanium Electrical Resistivity in Hydrogen and Deuterium. J Mater Eng Perform. 2022;31(6):4566-74. doi:10.1007/s11665-021-06542-1
17. Regulska K, Januszewicz B, Klimek L. Influence of Abrasive Treatment on a Transformation of Zirconium Oxide Used in Dental Prosthetics. Materials. 2022;15(12):4245. doi:10.3390/ma15124245
18. Spyrides GM, Spyrides SM, Resende TH, Mello EB, Neto OA, Ribeiro JB. Immediate implants placed into central incisors sockets: a surgical and prosthetic two years follow-up. Braz Dent Sci. 2017;20(4):130-7. doi:10.14295/bds.2017.v20i4.1401
19. Sharma P, Arora A, Valiathan A. Age Changes of Jaws and Soft Tissue Profile. The Scientific World Journal. 2014;301501. doi:10.1155/2014/301501
20. Mockutė G, Klimaitė G, Smailienė D. The Morphology of Impacted Maxillary Central Incisors: A Systematic Review. Medicina. 2022;58(4):462. doi:10.3390/medicina58040462
21. Fibryanto E, Tjin RR, Prahasti AE, Kusnoto J. Difference in Average Length of Maxillary Incisors between the Deuteromalayid and Protomalayid Sub-races. World J Dent. 2021;12(2):111-4. doi:10.5005/jp-journals-10015-1805
22. Silvester CM, Kullmer O, Hillson S. A dental revolution: The association between occlusion and chewing behaviour. PLOS ONE. 2021;16(12):e0261404. doi:10.1371/journal.pone.0261404
23. Zirakashvili N. Study of the stress–strain state of spongy bone around an implant under occlusal load. Math Mech Solids. 2020;25(5):1172-81. doi:10.1177/1081286520904978
24. Khandia R, Pandey MK, Khan AA, Rzhepakovsky IV, Gurjar P, Karobari MI. Codon Usage and Context Analysis of Genes Modulated during SARS-CoV-2 Infection and Dental Inflammation. Vaccines (Basel). 2022;10(11):1874. doi:10.3390/vaccines10111874
25. Alrakkad IA, Alrakkad RA, Altamimi MS, Alshammari NM, Alghuraymil AA, John MA, et al. Review on Dental Implant and Infection Management Approach. Arch Pharm Pract. 2022;13(1):37-9.
26. Yaghini J, Salmani SM, Hasheminejad SM, Mogharehabed A. Dentists’ Attention to Periodontal Therapy in the Patients Treatment Planning to Dental Clinics of Isfahan City. Arch Pharm Pract. 2022;13(2):51-6.
27. Remizova AA, Dzgoeva MG, Tingaeva YI, Hubulov SA, Gutnov VM, Bitarov PA. Tissue dental status and features of periodontal microcirculation in patients with new covid-19 coronavirus infection. Pharmacophore. 2021;12(2):6-13.
28. Al Zahrani TA, AlTulihe SB, Alabdullateef KS, Alanazi OS, Almutairi NM, Alshamrani HA, et al. Knowledge And Practices Of Dental Professionals Towards Lasers In Riyadh City: A Survey-Based Study. Pharmacophore. 2021;12(3):77-86.
29. Lee JH. Factors affecting the academic performance of low-and high-performing dental students: evidence from Japan. J Adv Pharm Educ Res. 2022;12(3):82-6.